Hybrid Chart Paper Recorder (RM10 series)

Description

RM10C is 100mm calibrated hybrid recorder in a 144x144 DIN front panel and short depth of 150mm case housing. The recorder has versatility of universal input and scale in wide ranges and flexibility of user-reconfigurable unique functions. RM10C offers 1, 2 continuous pen and 6 dot point models.

  • Short Case Depth and Light Weight
  • Dimension: 144W 144H 150D(mm)
  • Weight: 1.5kg
  • IP-65: Dust-Proof, Water-Proof
  • Highly Responsive and Multi-Functional
  • Universal Input and Range
  • RS-232C Provided as Standard
  • Without Transducer; RM10C (pH); Sensor Input: Glass or Metal Electrode (pH/ ORP)
  • UL, CSA, CE Approved

Title

Item

Pen Recorder

Multipoint Recorder

pH Recorder

Input

Measuring Point

1, 2

6

1

lnput sampling

125ms

10s/6ch

200ms

Type of input

Direct voltage: ±10mV, 0-20mV, 0-50mV, ±200mV, ±1V, ±10V, 0-5V

Glass electrode (pH or ORP)

Direct current: 4-20mA (250Ω shunt resistance required)

Temperature compensation

Thermocouple: K, T,J, E, B, S, R, C, N, PR40-20, PL 2, L, Au-Fe, U

350Ω, 500Ω, 1kΩ, 6.8kΩ, 10kΩ

RTD: Pt100, JPt100

 

Analog retransmission

   

4-20mA DC isolated, 550Ω max. (Option)

Performance

Accuracy

±(0.2% of rdg + 3digits) max. for Digital indicator

Linearity: ±0.03pH(±3mV) Repeatability: ±0.02pH(±1mV)

Noise reduction

NMRR: 60dB min. (50/60Hz±0.1Hz)

 

CMRR: 140dB min. (50/60Hz±0.1 Hz)

100dB min. (50/60Hz±0.1 Hz)

lsolation resistance

0.5kV DC 20MΩ min. between theeach terminal and grounding terminal

0.5kV DC 1000MΩ min. between R and G terminal and grounding terminal

Dielectric strength

1.5kV AC for 1 minute between power terminal and grounding terminal

 

0.5kV AC for 1 minute between input terminal and grounding terminal

Vibration resistance

10 to 60Hz 1m/s2 max.

Shock resistance

2m/s2 max.

Alarm

Nos.of relay outputs

3 outputs (Form A contacts: Built-in option)

6 outputs (Form A contacts: Built-in option)

2 Alarm outputs, 2 types(H,L) 1 Error output(Form A contacts: Built-in option)

Alarm output contact capacity

250V AC 3A max. .(resistance load), 30V DC 3A max. .(resistance load)

Communication

Interface

RS-232C (Standard), RS-485 (Option)

RS-232C (Standard)

Power supply

Rated power voltage

100-240VAC (50/60Hz)

Power consumption

30VA max.(100V AC)

25VA max.(100V AC)

20VA max.(100V AC)

Structure

3Mounting/Housing

Panel mounting / Front panel: Dust-proof, Water-proof (lP-65)

Dimension

144(W)x144(H)x150(D)mm

144(W)x144(H)x174(D)mm

Weight

1.8kg max.

1.5kg max.

2.0kg max.

Operation condition

Allowable conditions

Temperature: 0-50°C,Humidity: 20-80%RH

Recording /Printing Performance

Recording

Disposable felt pen

Wire dot 6 color ribbon

Disposable felt pen

Printing

Wire dot 1color ribbon

Wire dot 1color ribbon

95% step response time

1.0s max.

 

1.0s max.

Dot print interval

 

10.0s/6ch max.

 

Chart paper

Length: 16m,Width: 114mm,Width folded: 40mm

Chart speed

User-selective from 41 speeds

User-selective from 34 speeds

 

in range: 5-12000mm/hr.

in range: 0-1500mm/hr

in range: 5-12000mm/hr.

Recording/Printing Color

Red, Green

Purple, Red, Green, Blue, Brown, Black

Red

Safety Standard and EMI Standard

Electrical safety Emissions

lEC1010-1,UL3101-1 installation category 2 and pollution degree2. UL3101-1 EN55011 Group1 Class A

Immunity

EN50082-2

 

  • Furnace monitoring for steel and metal processing
  • Remote control monitoring & recording of pump station
  • Flow leakage detection and alarm generation
  • Batch production and quality analysis
  • Gas and effluent emission recorder
  • Sterilisation & ultraviolet Disinfection processes recorder
  • Milk pasteurisation & other HACCP applications
  • Cold storage and transport of perishable goods with Data logger (mini-log)
  • Pure water and waste water quality (pH, conductivity etc.) monitoring & recorder
  • Tank level monitoring and remote inventory management over modem
  • Limit overshoot, pump failure and event alarm over GSM messaging
  • Oven temperature monitoring & recording for Electric Manufacturing

Fields marked with * are mandatory fields.

:
:
:
:
:
:
:
:
: