Functions

Reference

Range

Resolution

Accuracy

Remark

DCV

50mV

-5.000mV~
55.000mV

1 µV

0.02+0.02

Input Resistance,
approximately 100MΩ

500mV

-50.00mV~
550.00mV

10 µV

0.02+0.01

OMH

500Ω

0.00Ω ~
550.00Ω

0.01Ω

0.05+0.02

Test current for 500Ω ;
Approximately 1mA;
Test current for 5KΩ ;
Approximately 0.1mA; Open circuit voltage;
Dose not include
lead resistance

5KΩ

0.0000KΩ ~
5.50000KΩ

0.1Ω

0.05+0.02

TC

R

0ºC~1767ºC

1ºC

0~500ºC : 1.8ºC

By using ITS-90
temperature scale;
The accuracy does not
include the error of
internal temperature
compensation caused
by a sensor;

S

0ºC~
1767ºC

500~
1767ºC : 1.5ºC

K

-100ºC~
1372.0ºC

0.1ºC

-100.0~
0.0ºC : 1.2ºC

0.0~
1372.0ºC : 0.8ºC

E

-50.0ºc~
850.0ºC

-50.0ºC~
0.0ºC : 0.9ºC

0.0ºC~
85.0ºC : 1.5ºC

J

-60.0ºC~
1120.0ºC

-60.0~
0.0ºC : 1.0ºC

0.0~
1120.0ºC : 0.7ºC

T

-100.0ºC~
400ºC

-100.0~
0.0ºC : 1.0ºC

0.0~
400.0ºC : 0.7ºC

N

-200.0ºC~
1300.0ºC

-200.0~
0.0ºC : 1.5ºC

0.0~
1300.0Vc : 0.9ºc

B

600ºc~
1820ºC

1ºC

600~
800ºC : 2.2ºC

800~
1000ºC : 1.8ºC

1000~
1820ºC : 1.4ºC

L

-60.0ºC~
900.0ºC

0.1ºC

-60.0~
0.0ºC : 0.7ºC

0.0~900.0ºC : 0.5ºC

U

-100.0V~
600.0ºC

0.1ºC

-100.0~
0.0ºC : 0.7ºC

0.0~
600.0ºC : 0.5ºC

RTD CONTINUITY

PT100 385

-200.0ºC~
800.0ºC

0.1ºC

-200.0~
0.0ºC

By using temperature
scale ITS-90. Does not
include lead resistance.
Assuming all three
RTD leads are
matched for 3-w
input

0.0~
400.0ºC : 0.7ºC

400.0~
800.0ºC : 0.8ºC

PT1000 385

-200.0ºC~
630.0ºC

-200.0~
100.0ºC : 0.3ºC

100.0~
300.0ºC : 0.5ºC

300.0~
630.0ºC : 0.7ºC

PT200 385

-200.0ºC~
630.0ºC

-200.0~
100.0ºC : 0.8ºC

100.0~
300ºC : 0.9ºC

300.0~
630ºC : 1.0ºC

PT500 385

-200.0ºC~
630ºC

-200.0~
100.0ºC : 0.4ºC

100.0~
300.0ºC : 0.5ºC

300.0~
630.0ºC : 0.7ºC

Cu10

-100.0ºC~
260.0ºC

1.8ºC

 

Cu50

-50.0ºC~
150.0ºC

0.7ºC

500Ω

≤50 Ω Sound

0.01Ω

 

500Ω Approximately
1mA test Current

DC Voltage

100mV

-10.000~
110.00mV

1µV

0.02+0.01

Maximum output:
0.5mV

1000mV

-100.00~
1100.00mV

10µV

0.02+0.01

Maximum output:
2mV

Resistance

400Ω

0.00Ω ~
400.00Ω

0.01Ω

0.02+0.02

Excitation current:
0.5-3 mA; if
0.1-0.5mA, add 0.1Ω
; Does not include
lead resistance;

4KΩ

0.0000kΩ ~
4.0000kΩ

0.1 Ω

0.05+0.025

Excitation current:
0.05-0.3 mA; Does not
include lead resistance;

TC

R

0ºC~
1767ºC

1ºC

0~
100ºC : 1.5ºC
100~
1767 : 1.2ºC

By using ITS-90
temperature scale; The accuracy
does not include the error
of internal temperature
compensation caused by a
sensor;

S

0ºC~
1767ºC

0~
100ºC: 1.5ºC

100~
1767ºC : 1.2ºC

K

-200.0ºC~
1372.0ºC

0.1ºC

-200.0~
100.0ºC : 0.6ºC

-100.0~
400.0ºC : 0.5ºC

400.0~
1200.0ºC : 0.7ºC

1200.0~
1372.0ºC : 0.9ºC

E

-200.0ºC~
1000.0ºC

-200.0~
-100.0ºC : 0.6ºC

-100.0~
600.0ºC : 0.5ºC

600.0~
1000.0ºC : 0.4ºC

J

-200.0ºC~
1200.0ºC

-200.0~
-100.0ºC: 0.6ºC

-100.0~
800.0ºC: 0.5ºC

800.0~
1200.0ºC: 0.7ºC

T

-250.0ºC~
400.0ºC

-250.0~
400.0ºC: 0.6ºC

N

-200.0ºC~
1300.0ºC

-200.0~
-100.0ºC: 1.0ºC

-100.0~
900.0ºC: 0.7ºC

900.0~
1300.0ºC: 0.8ºC

B

600ºC~
1820ºC

1ºC

600~
800ºC: 1.5ºC

800~
1820ºC: 1.1ºC

L

-200.0ºC~
900.0ºC

0.1ºC

-200.0~
0.0ºC: 0.7ºC

0.0~
900.0ºC: 0.5ºC

U

-200.0ºC~600.0ºC

-200.0~
0.0ºC: 0.7ºC

0.0~
600.0ºC: 0.5ºC

RTD

PT100
385

-200.0ºC~
800.0ºC

0.1ºC

-200.0~
0.0ºC: 0.3ºC

By using
temperature scale
ITS-90
Excitation current: ±0.5~±3mA
for Pt100, Cu10, Cu50, add
0.5 when excitation current
is ±0.1mA-0.5mA;
Excitation current: ±0.05mA~
±0.3mA for PT200, PT500,
PT1000;
Does not include lead resistance

0.0~
400.0ºC: 0.5ºC

400.0~
800.0ºC: 0.8ºC

PT1000

-200.0ºC~
630.0ºC

-200.0~
100.0ºC: 0.2ºC

100.0~
300.0ºC: 0.5ºC

300.0~
630.0ºC: 0.7ºC

PT200
385

-200.0ºC~
630.0ºC

-200.0~
100.0ºC: 0.8ºC

100.0~
300.0ºC: 0.9ºC

300.0~
630.0ºC: 1.0ºC

PT500
385

-200.0ºC~
630.0ºC

100.0~
300.0ºC: 0.5ºC

100.0~
300.0ºC: 0.5ºC

300.0~
630.0ºC: 0.7ºC

Cu10

-10.0ºC~
260.0ºC

-10.0~
260.0ºC: 1.8ºC

Cu50

-50.0ºC
150.0ºC

-50.0~
150.0ºC: 0.6ºC


 

All technicians responsible for Installing, Commissioning, Calibrating and Maintaining process industry transmitters, thermocouple including Instrument and Electric Shop Technicians, Instrument and Control Shop Technicians, Maintenance Technicians.