Precision Universal Calibrator IT52.54 |

- Process Meter is a hand held ,battery operated tool for measuring electrical parameters. It has all the features of a digital multimeter (besides the feature of RTD and Thermocouple),it could also output signals of direct voltage ,current,resistance,temperature and frequency as well.
- Auto/manual range switch,measure value display hold.
- The thermocouple high accuracy cold-end can autooffset
- Panel auto calibrated.
- The function of auto closed backlight and auto power off.
- Large LCD include white LED backlight.
- It is easy to operation by user.,cabinet and solid adapt to be used at locale.
Measuring Function |
Range |
Measuring range |
Resolving Power |
Accuracy |
Remark |
| DC Voltage |
+4V |
-4.000V~ 4.000V |
1mV |
0.2%+4 |
Input impedance: 10MΩ |
| 40V |
-40.00V~ 40.00V |
0.01V |
0.2%+4 |
| 400V |
-400.0V~ 400.0V |
0.1V |
0.2%+4 |
AC Voltage(40Hz~ 500 Hz)
(5%~ 100%RANG)
|
400mV |
0~ 400.0mV |
0.1mV |
1%+4 |
Input impedance : 10MΩ <100pF |
| 4V |
0~ 4.000V |
1mV |
0.5%+4 |
| 40V |
0~ 40.00V |
0.01V |
0.5%+4 |
| 400V |
0~ 400.0V |
0.1V |
0.5%+4 |
DC mV Voltage |
40mV |
-40.00mV~ 40.00mV |
0.01mV |
0.5%+6 |
Input impedance: 10MΩ |
| 400mV |
-400.0mV~ 400.0mV |
0.1mV |
0.2%+4 |
| OHM |
400OΩ |
0~ 400.0OΩ |
0.1OΩ |
0.2%+4Ω |
Plough voltage: 0.4V
Not including the accuracy of down-lead resistance |
| 4kΩ |
0~ 4.000kΩ |
1Ω |
0.2%+4 |
| 40kΩ |
0~ 40.00kΩ |
0.01kΩ |
0.2%+4 |
| 400kΩ |
0~ 400.0kΩ |
0.1kΩ |
0.2%+4 |
| 4MΩ |
0~ 4.000MΩ |
1kΩ |
0.2%+4 |
| 40MΩ |
0~ 40.00MΩ |
0.01MΩ |
1%+4 |
| DC Current |
40mA |
0-40.00mA~ 40.00mA |
0.01mA |
0.2%+4 |
Input impedance: 1Ω |
| 400mA |
400.00mA~ 400.00mA |
0.1mA |
0.2%+4 |
AC Current (40Hz~200Hz) (5%~100%) |
40mA |
0~ 40.00mA |
0.01mA |
0.5%+4 |
Input impedance: 1Ω |
| 400mA |
0~ 400.00mA |
0.1mA |
0.5%+4 |
| Frequency |
50Hz |
0~ 50.0 Hz |
0.01Hz |
0.1%+3 |
|
| 500Hz |
0~ 500.0 Hz |
0.1Hz |
0.1%+3 |
| 5kHz |
0~ 5.000 kHz |
1Hz |
0.1%+3 |
| 50kHz |
0~ 50.00k Hz |
0.01kHz |
0.1%+3 |
| 100kHz |
0~ 100.0k Hz |
0.1kHz |
0.1%+3 |
| Duty cycle |
0.1%~9 9% |
|
0.1% |
1% |
|
| Diode test |
1V |
|
0.001V |
10% |
Plough voltage 1.1V~1.6V
Current of short circuit : 0.6mA
|
Continuity test |
=50OBB |
|
0.1Ω |
|
| Thermocouple |
R |
-40°C~ 1760°C |
1°C |
0.5%+3(=100°C)
0.5%+2( >100°C)
|
Adopt Its -90thermometric scale not including the accuracy of RJC error RJC error:±20 |
| S |
-20°C~ 1760°C |
| K |
-200°C~ 950°C |
1°C |
0.2%+2(=100°C)
0.5%+1( >100°C)
|
| E |
-200°C~500°C |
| J |
-200°C~700°C |
| T |
-200°C~400°C |
| N |
-200°C~ 1000°C |
1°C |
0.5%+3(400~ 600°C0.5% +2( >600°C) |
| B |
400°C~ 1800°C |
Thermo resistance |
Pt100 |
-200°C~ 700°C |
1°C |
0.5%+2 |
Pt100-385 thermometric scale, not including the accuracy of down-lead resistance |
| Cu50 |
-50°C~ 150°C |
1°C |
0.5%+4 |
| Output Function |
| Function |
Range |
Setting range |
Resolving power |
Accuracy |
Remark |
| OMH |
400Ω |
0~ 400.0Ω |
0.1Ω |
0.2%+4 |
±1mA current not including the accuracy of down-lead resistance |
| 79 DC mV |
100mV |
-10.00mV~ 110.00mV |
0.01mV |
0.2%+4 |
Max. output current: 5mA |
| DCV |
5V |
-0.5000V~ 5.5000V |
1mV |
0.2%+4 |
Max. output current: 5mA |
| FREQ |
100Hz |
1.0Hz~ 110.0Hz |
0.1Hz |
0.2%+2 |
50% duty cycle 5VP-P |
| 1kHz |
0.100kHz~ 1.100kHz |
0.001kHz |
0.2%+2 |
| 10kHz |
1.0kHz~ 11.0kHz |
0.1kHz |
0.2%+2 |
Analog transducer XMT |
-20mA |
0~ -22.000mA |
0.01mA |
0.2%+4 |
Outside power supply 28V when 20mA 1kΩ resistance |
| DC mA |
20mA |
0~ 22.000mA |
0.01mA |
0.2%+4 |
Outside power supply 15V when 20mA 400Ω resistance |
| Thermocouple |
R |
-40°C~ 1760°C |
1°C |
0.2%+3
(=100°C)
0.5%+2
( >100°C)
|
Adopt Its- 90thermometric scale not including the accuracy of RJC error RJC error:±2°C |
| S |
-20°C~ 1760°C |
0.1°C |
| K |
-200°C~ 1370°C |
0.5%+20
(=-100°C)
0.5%+10
( >-100° )
|
| E |
-200°C~ 1000°C |
| J |
-200°C~ 1200°C |
| T |
-200°C~ 400°C |
| N |
-200°C~ 1300°C |
| B |
400°C~ 1800°C |
1°C |
0.5%+3
(400~600°C)
0.5%+2
( >600°C)
|
Thermo resistance |
Pt100 |
200°C~ 850°C |
0.1°C |
0.2%+6 |
Pt100-385 thermometric scale ±0.1mA not including the accuracy of down-lead resistance |
| Cu50 |