Universal Desktop Calibrator
IT 53 Series of digital test sets and calibrators are portable instruments suitable for measurement & calibration in instrumentation departments, scientific laboratories, research institutions & for field applications. Three models are available to cater to various demands and measurement requirements.

While IT53.01 and IT53.02 are 3½ digit Calibrators, 1T'53.03 is a 4½ digit Calibrator. All the models have reference junction compensation, built in for J, K, R & S thermocouple output measurement and are capable of simulation of thermocouple output for calibration and testing of temperature instruments. IT53.01 and IT53.03 are also provided with measurement and simulation of Pt 100 DIN RTD.


For measurement & simulation of DC current, IT 53.02 & IT 53.03 are ideally suitable. IT53.
  • Features
  • Specification
  • Application
  • Brochure/Documents

  • Measurement & Simulation of Thermocouple outputs, mV, current and Resistance Thermometers
  • Both 3½ digit & 4½ digit models
  • 0.5" bright seven segment LED display
  • Built-in reference junction compensation
  • Lead balancing resistance provided
  • Comparison of two thermocouples

SPECIFICATIONS TEMP.TEST SET IT53.01 CALIBRATOR IT53.02 CALIBRATOR IT53.03
Accuracy at 27 ºC ± 0.1%FS ± 1 count ± 0.1%FS ± 1 count ± 0.1%FS ± 2 counts
Influence of temperature ± 0.1%FS ± 1 count per 10 ºC (15…45 ºC) ± 0.1%FS ± 1 count per 10 ºC (15…45 ºC) ± 0.1%FS ± 1 count per 10 ºC (15…45 ºC)
Display 3½ digit 3½ digit 4½ digit
D.C. Voltage Measurement Resolution Input Impedance 0-199.9 mV 100 µV > 1000M Ohms 0-19.99 & 0-199.9 mV 10 & 100 µV >1M Ohms in 20 mV range >1000M Ohms in 200 mV range 0-199.99 mV 10 µV >1M Ohm
Direct Current Measurement Resolution Input Impedance   0-1.999 & 0-19.99 mA 1 & 10 µA 100 & 10 Ohms 0-19.999 mA 1µA 10 Ohms
Resistance Measurement Resolution Current through unknown resistor 0-199.9 Ohms 0.1 Ohm 1 mA   0-1999.9 Ohms 0.1 Ohm 1 mA
Thermocouple Output Measurement Range Reference Temperature Accuracy of Compensation (at 27 ºC) 0-199.9 mV 0 ºC ± 2 ºC ± 1 count 0-19.99 & 0-199.9 mV 0 ºC ± 1 ºC ± 1 count 0-199.99 mV 0 ºC ± 1 ºC ± 2 counts
Two Thermocouples Comparison Thermocouple Reference Fe/ Ko, Chromel /Alumel or NiCr/Ni DIN & Pt Rh /Pt By Switch on two sets of terminals Built - in Fe/ Ko, Chromel /Alumel or NiCr/Ni DIN & Pt Rh /Pt By Switch on two sets of terminals Built - in Fe/ Ko, Chromel /Alumel or NiCr/Ni DIN & Pt Rh /Pt By Switch on two sets of terminals Built - in
Voltage Source Resolution Max. Current Capacity Lead Balancing Controls   0-1.999 & 0-19.99mA 1 & 10µV 0-19.999mA 1µV
Resistance Source Selection Pt.100DIN for -40ºC, -20ºC,0ºC,20ºC,40ºC, 60ºC,100ºC,150ºC, 200ºC,300ºC,400ºC,& 500ºC   1 to 1000 ohms
Max. permissible current By rotary selector switch 10mA   Steplessly adjustable
Output display Dial below selector Switch   10mA Press to read on built-in DPM
Lead Balancing Resistance For use in mV & Resistance source mode For mV Source For mV Source
Range 0,1,2,3,4,5,10,20,30, 40 & 50 ohms 0-99 ohms in steps of 1 ohms 0-99 ohms in steps of 1 ohms
Selection By Rotary Switch By Thunderwheel Switch By Thunderwheel Switch
Accuracy ±0.5% ±1% ±1%
Power 230V AC ±10%,50Hz,6VA (Approx.) 230V AC ±10%,50Hz,6VA (Approx.) 230V AC ±10%,50Hz,6VA (Approx.)
Size 300X100X175 mm 300X100X175 mm 300X100X175 mm
Weight 1.5 Kgs.(Approx.) 1.5 Kgs.(Approx.) 1.5 Kgs.(Approx.)

  • Suitable for measurement & calibration in instrumentation departments, scientific laboratories, research institutions & for field applications.
  • Used for simulation of thermocouple output for calibration and testing of temperature instruments.
  • Also provided with measurement and simulation of Pt 100 DIN RTD.
 
 